KASON HPT-10M Series is mainly used to determine the resistance of thermoplastics pipes and short-time hydraulic burst to constant internal water pressure at constant temperature.It is applicable to thermoplastic pipes intended for the conveyance of fluids.
Request a QuoteIt confirms to ASTM D1598-2004, ASTM D1599 and IS0 1167-2006.
ASTM D1598-2004 Standard Test Method for Time-to-Failure of Plastic Pipe under Constant
Control System: Two-Level Monitoring, Multi-Purpose, Premium Quality, Two-Level Computer, Versatile Functions, User-Friendly
Two-Level Monitoring:
The first level provides 1-hour, detailed monitoring; the second level provides comprehensive, full-process monitoring.
Multi-Purpose:
The system can perform both long-term hydrostatic testing and burst testing (each channel can perform both burst and pressure testing). It can also be used for thermal stability testing of thermoplastic pipes under hydrostatic conditions. Each channel features multi-stage accumulator pressure control, ensuring fast response and stable pressure control, extending the life of the solenoid valve.
Premium Quality:
The entire system utilizes EU CE-certified electronic components, German solenoid valves, and GEMS sensors. It offers exceptional long-term reliability and stability, high efficiency for multi-port testing, and low operating costs for excellent economic benefits.
Wide Pressure Control Range:
Each station's pressure control range is adjustable from 0.4 to 10 MPa.
High Reliability:
All electrical, mechanical, and software components are equipped with safety features. Such features include protection against electrical leakage, short circuits, static electricity, interference, seepage, power outages, and misoperation.
Integrated modular assembly
Equipped with a microcontroller and solenoid valve control unit, they prevent interference. A rupture in one test line does not affect the normal testing of other lines.
Comprehensive functions
Features rupture detection, real-time monitoring, power-off pressure maintenance, power-off protection, and automatic power-on recovery. Test results can be printed and stored. The main control interface is available in Chinese and English, providing multi-language operation.
User-friendly design
An automatic or manual time increment option is available to compensate for ineffective time during the test. The test pressure can also be adjusted mid-test to meet specific pipe testing requirements.
The product consists of two components:
The pressure main unit and the constant temperature water tank, all integrated with the fixture. The specifications and models of each component are user-selectable, and custom designs are also available.
|
Model |
KASON HPT-10M |
|
Pressure range |
0.4~10MPa |
|
Test station |
3 stations or even more |
|
Control mode |
Computer |
|
Resolution of pressure testing |
0.001MPa |
|
Accuracy |
Better than ±1% |
|
Testing pipe |
φ20mm-φ630mm |
|
Constant temp. range |
room temperature~95℃ |
|
Temp. Accuracy |
±0.1℃ |
|
Time accuracy |
≤±0.1% |
|
Power supply |
3 phase, 380v±10%, 50Hz |
| Name | Download |
|---|