KASON-VBT50KV Computer Controlled Voltage Breakdown Testers

KASON-VBT50KV Computer Controlled Voltage Breakdown Testers

It is mainly applicable to the test of breakdown strength and withstand voltage time of solid insulating materials such as wire sleeve, resin and adhesive, impregnated fiber products, mica and its products, plastics, film composite products, ceramics and glass under power frequency voltage or DC voltage;

Request a Quote

Product Details

Ⅰ.Manufacturing and inspection standards of products

1、GB/T1408.1-2006《Electrical strength of insulating materials -Test methods - Part 1:Tests at power frequencies》

2、GB/T1408.2-2016《Insulating materials -Test methods for electric strength -Part 2:Additional requirements for tests using direct voltage》

3、JJG 795-2004 《Withstanding Voltage Testers》

Ⅱ.Applicable test method standards

1、GB/T1695-2005《Rubber, vulcanized -  Determination of electrical breakdown strength and voltage resistant at commercial power frequency》

2、GB/T3333-1999《Cable paper—Determination of electrical strength at power frequence》

3、GB/T12913-2008《Capacitor tissue paper》

4、ASTM D149《Test method for dielectric breakdown voltage and dielectric strength of solid electrical insulating materials at commercial power frequencies》

Ⅲ. Scope of application

It is mainly applicable to the test of breakdown strength and withstand voltage time of solid insulating materials such as wire sleeve, resin and adhesive, impregnated fiber products, mica and its products, plastics, film composite products, ceramics and glass under power frequency voltage or DC voltage; The instrument is controlled by computer, which can quickly and accurately collect and process various data in the test process, and can access, display and print.

Product parameters

NO.

Model(KASON-VBT)

5KV

10KV

20KV

50KV

100KV

150KV

1

Input voltage

AC220V   50Hz

AC

380V 50hz

2

Output voltage

AC:0~5kV

DC:0~5kV

AC:0~10kV;DC:0~10kV

AC:0~20kV

DC:0~20kV

AC:0~50kV

DC:0~50kV

AC:0~100kV

DC:0~100kV

AC:0~150kV;DC:0~150kV

3

Output power

0.5kVA

1kVA

2kVA

3kVA

10kVA

15kVA

4

Measuring range

AC:0.05~5kVDC:0~5kV

AC:0.5~10kVDC:0.5~10kV

AC:1~20kV DC:1~20kV

AC:1~50kV

DC:1~50kV

AC:1~100kV

DC:1~100kV

AC;1~150kV;DC;1~150kV

5

Measuringerror

≤ 2%

6

Boosting rate

0.025kV/s~0.5/s

0.05kV/s~1kV/s


0.1kV/s~2kV/s



0.20kV/s~5kV/s



0.5kV/s~10kV/s



1kV/s~10kV/s


7

Withstand voltagetime

0~4H (No load withstand voltage)

8

Leakage current

1~30 mA Can be freely set by computer software

9

Powersupply

Single phase AC voltage of 220V ± 10% and frequency of 50Hz ± 1%

10

Test environment temperature

Temperature:15 ~ 30 ℃;relative humidity: 30% ~ 65%, which can operate stably


11

Overall dimension


L*W*H: 920*530*1090mm

1620*1110*1510 mm

2100*1410*2100mm


Packed dimension


L*W*H:1200*940*1410mm

1740*1290*1820mm

2230*1630*2370mm

12

Equipment weight


107kg

330kg

670kg


Packed weight


167kg

420kg

770kg

13

Grounding requirement

The instrument needs tobe grounded separately.The groundingmust comply with national standards: the metal rod should be buried at least 1.5m underground.


Download

Name Download

Applicable industries

Related Products

Leave a message