The KASON-BFT200 Ampoule Breaking Force Tester is suitable for testing the breaking force required to separate the neck from the body of pharmaceutical injection ampoules.
Request a QuoteProduct Application:
The KASON-BFT200 Ampoule Breaking Force Tester is suitable for testing the breaking force required to separate the neck from the body of pharmaceutical injection ampoules. It is a commonly used testing device for pharmaceutical companies, ampoule manufacturers, and drug testing institutions, and complies with the testing requirements for breaking force tests of ampoule pharmaceutical packaging as specified in standards such as GB2637, YBB00332002, YBB 00042012, YBB 00052012, ISO9187.1, and ISO9187.2.
Test Principle:
The sample is mounted on two V-shaped supports with a specified span. A pressure head of a specified size forces the ampoule neck to break apart from the body. The instrument measures the change in force during this process, thus obtaining the corresponding breaking force value.
Product Features:
Ø Industrial-grade 8-inch touchscreen with a menu-driven, flat design eliminates the need for complex operations; one-click access to any interface is sufficient.
Ø Dedicated, independently developed single-chip microcomputer-based measurement and control system for breaking force, offering simple, fast, stable, reliable, and high-precision operation.
Ø The measurement and control software complies with pharmaceutical GMP access control requirements, facilitating standardized equipment use by laboratory management personnel.
Ø The test curve displays the trend of experimental data changes in real time throughout the entire test process.
Ø High-precision motors combined with a lead screw and guide rail structure ensure lifting accuracy and improve test precision.
Ø Equipped with a high-precision force acquisition sensor, it can accurately measure the maximum force at the moment of breakage.
Ø The instrument has functions such as test result judgment, storage, query, and data traceability.
Ø Uses internationally renowned brand components, ensuring stable and reliable performance.
Ø The instrument features power-off storage, overload alarm, and shutdown protection functions to ensure test safety.
Ø Equipped with a USB universal data interface for convenient data output and transmission.
Ø Equipped with a mini printer for direct printing of test results.
Ø Multi-level user access control with password login facilitates user management.|
Specifications |
Specifications |
|
Model |
KASON-BFT200 |
|
Sample Type |
Suitable for 1~25mL |
|
Span Specification |
36, 60mm |
|
Force Range |
0~200N |
|
Force Resolution |
0.001N |
|
Force Accuracy |
±0.5% |
|
Standard Test Speed |
10mm/min |
|
Speed Range |
0.1-500mm/min |
|
Speed Accuracy |
±0.5% |
|
Displacement Resolution |
0.001mm |
|
Displacement Accuracy |
±0.5% |
|
Control Method |
Microcomputer controlled, fully automatic testing |
|
Display Method |
8-inch LCD touchscreen |
|
Data Storage |
200 memory entries/USB export |
|
System Language |
Simplified Chinese/English |
|
Data Output Method |
Thermal or dot matrix printers available |
|
Dimensions |
400(L)×300(B)×450(H)mm |
|
Power Supply |
220VAC 50Hz / 110VAC 60Hz |
|
Net Weight |
Approx. 18KG |
|
※ Note: For users with special needs, our company can provide customized production within our capabilities to meet their requirements. |
|