It is mainly applicable to the test of breakdown strength and withstand voltage time of solid insulating materials such as wire sleeve, resin and adhesive, impregnated fiber products, mica and its products, plastics, film composite products, ceramics and glass under power frequency voltage or DC voltage;
Request a Quote|
NO. |
Model(KASON-VBT) |
5KV |
10KV |
20KV |
50KV |
100KV |
150KV |
|||
|
1 |
Input voltage |
AC220V 50Hz |
AC 380V 50hz |
|||||||
|
2 |
Output voltage |
AC:0~5kV DC:0~5kV |
AC:0~10kV; DC:0~10kV |
AC:0~20kV DC:0~20kV |
AC:0~50kV DC:0~50kV |
AC:0~100kV DC:0~100kV |
AC:0~150kV; DC:0~150kV |
|||
|
3 |
Output power |
0.5kVA |
1kVA |
2kVA |
3kVA |
10kVA |
15kVA |
|||
|
4 |
Measuring range |
AC:0.05~5kVDC:0~5kV |
AC:1~20kV DC:1~20kV |
AC:1~50kV DC:1~50kV |
AC:1~100kV DC:1~100kV |
AC;1~150kV; DC;1~150kV |
||||
|
5 |
Measuring error |
≤ 2% |
||||||||
|
6 |
Boosting rate |
0.025kV/s~0.5/s |
0.05kV/s~1kV/s |
0.1kV/s~2kV/s
|
0.20kV/s~5kV/s
|
0.5kV/s~10kV/s
|
1kV/s~10kV/s
|
|||
|
7 |
Withstand voltage time |
0~4H (No load withstand voltage) |
||||||||
|
8 |
Leakage current |
1~30 mA Can be freely set by computer software |
||||||||
|
9 |
Power supply |
Single phase AC voltage of 220V ± 10% and frequency of 50Hz ± 1% |
||||||||
|
10 |
Test environment temperature |
Temperature:15 ~ 30 ℃; relative humidity: 30% ~ 65%, which can operate stably
|
||||||||
|
11 |
Overall dimension |
|
L*W*H: 920*530*1090mm |
1620*1110*1510 mm |
2100*1410*2100mm |
|||||
|
|
Packed dimension |
|
L*W*H:1200*940*1410mm |
1740*1290*1820mm |
2230*1630*2370mm |
|||||
|
12 |
Equipment weight |
|
107kg |
330kg |
670kg |
|||||
|
|
Packed weight |
|
167kg |
420kg |
770kg |
|||||
|
13 |
Grounding requirement |
The instrument needs to be grounded separately.The grounding must comply with national standards: the metal rod should be buried at least 1.5m underground.
|
||||||||