KASON HTV-5/10/30/50 Manual Turret Vickers Hardness Tester with Weight Loading

KASON HTV-5/10/30/50 Manual Turret Vickers Hardness Tester with Weight Loading

The optical system designed by senior optical engineers not only has clear images, but also can be used as a simple microscope with adjustable brightness and comfortable vision.

Request a Quote

Product Details

Application:
1. Steel, nonferrous metals, tinsel, cemented carbide, sheet metal, metallographic structure.
2. Carburization, nitriding and decarburization layer, surface hardening layer, galvanized coating, coating.
3. Glass, chip and ceramic material.

Features:
1.The optical system designed by senior optical engineers not only has clear images, but also can be used as a simple microscope with adjustable brightness and comfortable vision.
2.Cast aluminum shell molding, the structure is stable and not deformed, high-grade automotive paint, anti-scratch ability, use for many years is still bright as new;
3.Display screen can directly show the hardness value, hardness conversion, test method, test force, holding time, measurement times, and the test process, and the test process can be intuitively understood.

Product parameters

MODEL

KASON HTV-5

KASON HTV-10

KASON HTV-30

KASON HTV-50

Turret Type

Manual

Test force

0.3kg(2.94N), 0.5kg(4.9N), 1kg(9.8N), 2kg(19.6N), 3kg(29.4N),

5kg (49N)

0.3kg(2.94N), 0.5kg (4.9N), 1kg (9.8N), 3kg (29.4N), 5kg (49N), 10kg (98N)

1kg(9.8N),3kg(29.4N),

5kg(49N),10kg(98N),

20kg(196N),30kg(294N)

1kg(9.8N), 5kg(49N), 10kg(98N), 20kg(196N), 30kg(294N), 50kg(490N)

Display

5-digit hardness value, 4-digit diagonal length (D1, D2),

Hold time, number of tests, average, standard deviation, return

Loading control

Automatic (load/hold/uninstall)

Optical channel

Dual light channels (eyepieces and CCD camera channels)

Test force retention time

1~99s

Optical system

Objectives: 10×, 20× (Customizable 40X)

Eyepiece Magnification: 10×

Total magnification (μm): 100×, 200×

Measuring range: 800μm 400μm

Resolution :1μm 0.5μm

Hardness test range

(5-3000)HV

XY test stand (optional)

Size (mm): 100×100

Travel range (mm): 25×25

Minimum reading (mm): 0.01

Maximum height of specimen

165mm

Maximum width of specimen

130 mm(distance from the center line of the indenter to the wall)

Meet standard

GBT4340.1, GBT4340.2, ASTM_E92

Conversion of hardness value

According to GB1172-99 and ASTM standards

Accuracy

JJG260-91, JIS B 7725, JIS B 7734, GB/ T3430.2-99, ASTM E-384, ISO 6507-2, ISO4546

voltage

AC220V/50HZ

Overall size

585×200×630mm

Weight

42kg


Download

Name Download

Applicable industries

Related Products

Leave a message