The optical system designed by senior optical engineers not only has clear images, but also can be used as a simple microscope with adjustable brightness and comfortable vision.
Request a Quote|
MODEL |
KASON HTV-5ADC |
KASON HTV-10ADC |
KASON HTV-30ADC |
KASON HTV-50ADC |
|
Turret Type |
Automatic |
|||
|
Test force |
0.3kg(2.94N), 0.5kg(4.9N), 1kg(9.8N), 2kg(19.6N), 3kg(29.4N), 5kg (49N) |
0.3kg(2.94N), 0.5kg (4.9N), 1kg (9.8N), 3kg (29.4N), 5kg (49N), 10kg (98N) |
1kg(9.8N),3kg(29.4N), 5kg(49N),10kg(98N), 20kg(196N),30kg(294N) |
1kg(9.8N), 5kg(49N), 10kg(98N), 20kg(196N), 30kg(294N), 50kg(490N) |
|
Display |
5-digit hardness value, 4-digit diagonal length (D1, D2), Hold time, number of tests, average value, etc. |
|||
|
Loading control |
Automatic (load/hold/uninstall) |
|||
|
Test mode |
HV/HK |
|||
|
Objective lens and indenter switch |
Automatic switching |
|||
|
Measurement multiple |
Both 10X and 40X can be measured with double objective lens |
|||
|
Optical channel |
Dual optical channel (eyepiece and CCD camera channel) |
|||
|
Test force retention time |
1~99s |
|||
|
Optical system |
Objectives: 10×, 20× (Customizable 40X) Eyepiece Magnification: 10× (optional 15X) Total magnification (μm): 100×, 400× Measuring range: 800μm 400μm Resolution :1μm 0.5μm |
|||
|
Hardness test range |
(5-3000)HV |
|||
|
XY test stand (optional) |
Size (mm): 100×100 Travel range (mm): 25×25 Minimum reading (mm): 0.01 |
|||
|
Maximum height of specimen |
165mm |
|||
|
Maximum width of specimen |
130 mm(distance from the center line of the indenter to the wall) |
|||
|
Meet standard |
GBT4340.1, GBT4340.2, ASTM_E92 |
|||
|
Conversion of hardness value |
According to GB1172-99 and ASTM standards |
|||
|
Accuracy |
JJG260-91, JIS B 7725, JIS B 7734, GB/ T3430.2-99, ASTM E-384, ISO 6507-2, ISO4546 |
|||
|
voltage |
AC220V/50HZ |
|||
|
Overall size |
585×200×630mm |
|||
|
Weight |
42kg |
|||